Troubleshooting kinked introducer

May 4, 2019 by Josh Farkas 1 Comment

introduction / clinical conundrum

A patient is admitted in transfer with a kinked introducer catheter (shown above).  Does this catheter need to be removed, or can it somehow be salvaged?

a few words about the introducer catheter

Introducer catheters (a.k.a. “cordis” or “sheath”) are traditionally popular for hemorrhagic shock resuscitation because they have high flow rates.  However I dislike this practice for several reasons:

  • They get kinked a lot.
  • In order to avoid getting kinked you need to put another catheter through them – which decreases the flow rate (thus defeating the whole purpose of using the catheter in the first place).
  • They aren't designed as a stand-alone device, so when inserted alone they have sharp-ish ends which could irritate vessels and possibly increase the risk of thrombus formation.
  • They don't get you a lot of lumens.
  • They aren't pressure injectable, so you can't use them for a CT angiography (seriously – this may cause the catheter to explode).

Personally I'd prefer *any* of the following options for large-volume resuscitation:

  • Hemodialysis catheter (two lumens, never kinks, enormous flow rate).
  • Trialysis catheter (hemodialysis catheter with three lumens)
  • MAC introducer.
  • Quad lumen pressure-injectable central line (generally disparaged for large-volume resuscitation, but if you attach this to a Belmont or Level-1 infuser you can get very decent flow rates).

how to salvage a kinked introducer catheter?

This can be fixed by gently advancing a catheter within the introducer to stent open the introducer (eliminating the kink).  The key here is that the inner catheter should be specifically designed for this use, so that it fits snugly within introducer (example above).   If you use a standard central line, there may be a gap between  the central line and the introducer which leads to bleeding or air embolism.  A specifically designed inner catheter also locks onto the introducer, keeping things clean.  If the inner catheter doesn't pass with gentle pressure, don't force it – in that case you may need to just remove the introducer.

This solution won't get you a large-bore lumen for rapid infusion of volume.  So if your patient is dying of hemorrhagic shock this won't get you out of trouble.  However, if you just need vascular access, then this is a simple way to convert a kinked (and totally worthless) introducer into a triple-lumen catheter.

  • Using an introducer catheter for access to provide large volume resuscitation probably isn't a great idea – there are numerous better options.
  • Don't stick any old central line into an introducer catheter – if there is any gap between the central line and the introducer this will cause bleeding and/or air embolism.
  • A kinked introducer can be salvaged by gently advancing a specially designed inner catheter through it, to stent it open.

 

Filed Under: EMCrit

Cite this post as:

Josh Farkas. Troubleshooting kinked introducer. EMCrit Blog. Published on May 4, 2019. Accessed on December 10th 2025. Available at [https://cmefix.emcrit.org/squirt/introducer/ ].

Financial Disclosures:

The course director, Dr. Scott D. Weingart MD FCCM, reports no relevant financial relationships with ineligible companies. This episode’s speaker(s) report no relevant financial relationships with ineligible companies unless listed above.

CME Review

Original Release: May 4, 2019
Date of Most Recent Review: Jul 1, 2024
Termination Date: Jul 1, 2027

You finished the 'cast,
Now Join EMCrit!

As a member, you can...

  • Get CME hours
  • Get the On Deeper Reflection Podcast
  • Support the show
  • Write it off on your taxes or get reimbursed by your department

Join Now!


.

Get the EMCrit Newsletter

If you enjoyed this post, you will almost certainly enjoy our others. Subscribe to our email list to keep informed on all of the Resuscitation and Critical Care goodness.

This Post was by the EMCrit Crew, published 7 years ago. We never spam; we hate spammers! Spammers probably work for the Joint Commission.

Comments

  1. Michal Dworzynski says

    May 5, 2019 at 07:01

    I think the easy way to avoid problems with high flow catheters (dialysis cath/sheaths / HF caths) is to insert them throught the internal carotid vein or iliac vein, there is also alternative – RIC peripheral iv may give a flow Of 750ml / min when attached to Belmount

    Reply

Leave a Reply Cancel reply

Your email address will not be published. Required fields are marked *

EMCrit Join

Login

  • Lost Password
Metasin Coaching
RCM
SCRAM Bag

CME Information

See our CME Information

Other Stuff

  • Have a great idea for the next podcast? Share it here!
  • Tough Questions. Maybe you have an answer!
  • When you're done listening to the podcast,
    check out these great sites.

Who We Are

We are the EMCrit Project, a team of independent medical bloggers and podcasters joined together by our common love of cutting-edge care, iconoclastic ramblings, and FOAM.

Resus Leadership Academy

Subscribe by Email

EMCrit is a trademark of Metasin LLC. Copyright 2009-. This site represents our opinions only. See our full disclaimer, our privacy policy, commenting policy and here for credits and attribution.